Have a look at new page for LabVIEW and GenAI coding projects...
Senior Scientist | R&D expert | Electrochemistry, Hydrogen, Fuel Cells & Electrolyzers | Electrocatalyst Innovation | LabVIEW | Lab Automation | In-/Ex-situ Characterization | Analytical Technologies | Mentor | GenAI
Saturday, July 4, 2026
Saturday, May 31, 2025
Analyst Assistant 4.0: Automating Elemental Analysis Workflows with LabVIEW
For analytical researchers handling high-throughput experimentation, data generation is rarely the bottleneck—data processing is. Elemental characterization techniques generate massive volumes of raw data across various file formats (.txt, .csv, .xlsx). Manually importing, filtering, trimming, and analyzing this data to meet standard operating procedures (SOPs) is not only tedious but also introduces the risk of human error.
To bridge the gap between raw instrument output and final reporting, I developed Analyst Assistant 4.0 (AA 4.0), a comprehensive MS Windows® application programmed entirely in LabVIEW®.
The Core Objective
The primary goal of AA 4.0 is to completely automate the data analysis and processing workflow for elemental characterization while keeping the analytical chemist in control of critical decision points.
Seamless Instrument Integration
To accommodate a versatile analytical laboratory, the AA 4.0 Launcher features a built-in sequence initiator tailored for a wide range of industry-standard instruments. The software natively processes data generated by:
Agilent ICP-MS: (Models 8900, 7900)
Spectro ICP-OES: (Models FHS22, FHD32)
Panalytical WD-XRF: (Zetium 4.0)
Elementar CHNSO: (Configured for CH, CHN, and CHN+O)
TE Instruments TNTS: (Xplorer)PAC TNTS: (Multitek and Elements)
The Human-in-the-Loop Workflow
Automation shouldn't mean a black box. Every sequence initiated in AA 4.0 rigorously follows established SOPs for data processing, but it is deliberately designed to prompt the analyst for inputs at critical stages.
Automated Ingestion: The program automatically imports the raw data files, regardless of whether the instrument exports in text, CSV, or Excel formats.Pre-Processing: The software handles the routine filtering, trimming, and baseline calculations automatically.Analyst Intervention: The software pauses to allow the analyst to visually inspect the data and manually choose critical points for final processing and analysis, ensuring expert oversight remains part of the calculation.
Traceability and LIMS Integration
In any certified laboratory environment, data traceability is just as important as the result itself. Upon completion of the processing sequence, AA 4.0 automatically generates two distinct deliverables:
The LIMS-Ready Template: The processed results are mapped directly into a standardized template, ready for immediate import into the Laboratory Information Management System (LIMS). This eliminates manual data entry and transcription errors.The Meta-Data File: A separate Excel workbook is generated to serve as a comprehensive audit trail. This file contains distinct sheets detailing every single step, formula, and analyst input used to process the data from the raw import to the final report, ensuring 100% traceability.
By automating the routine data-handling tasks while preserving analyst input for critical evaluations, Analyst Assistant 4.0 drastically accelerates turnaround times and ensures pristine data integrity across multiple elemental characterization platforms.
Wednesday, December 25, 2024
Moisture data logging with Custom Automation - DewPt 2.5 - LabVIEW based application
In the continuous effort to streamline laboratory operations and reduce manual intervention in experimental workflows, automating long-term setups is a practical necessity. To address this bottleneck, a custom physical reactor paired with a dedicated software application was developed for data collection processes.
Designing the Physical Architecture and Instrumentation
To monitor the internal environment, the system utilized an industrial dew point sensor. This sensor was biased using a 24V power supply (PS). The physical architecture for data acquisition was realized and tested through two distinct setups:
Setup 1 (Industrial Grade): The sensor's analog current output was routed through a current module in a CompactDAQ (cDAQ) system, which then logged
the data to a PC. This robust configuration proved highly reliable, operating continuously for days to weeks without any interruption. Setup 2 (Lab Scale): The sensor's generated current signal (up
to 20 mA) was continuously measured using a high-precision Keithley DAQ6510 multimeter equipped with a 7700 multiplexer module before logging to a PC. While effective for shorter runs, this lab-scale setup experienced random interruptions after 1 to 2 days of continuous measurement.
DewPt 2.5: The LabVIEW-Driven Core
To bridge the hardware control and data logging, a custom Windows application, DewPt 2.5, was programmed in LabVIEW. This software was designed to interface with the data acquisition hardware to facilitate continuous data collection and monitoring.
The application handles several critical operational tasks:
Continuous Automated Logging: Executes automated cycling processes, capturing moisture values consistently over extended timeframes.
Real-Time Signal Processing: Captures the incoming current measurements (mA) from the acquisition hardware on the fly and converts them into the desired moisture measurement units.
Dynamic Visualization: Computes and graphs these derived values in real-time within the UI, allowing the user to visually estimate the precise cut-off points for the experiments.
Automated Archiving: Saves all raw current data and computed moisture values directly to local files for downstream analysis.
Sunday, February 20, 2022
Motivation for LabVIEW for researchers - Introduction to LabVIEW
- Introduction to LabVIEW for scientists/engineers - for newbies!